R系列旋轉蒸發(fā)器工作原理
通過電子控制,使燒瓶在最適合速度下,恒速旋轉以增大蒸發(fā)面積。通過真空泵使蒸發(fā)燒瓶處于負壓狀態(tài)。蒸發(fā)燒瓶在旋轉同時置于水浴鍋中恒溫加熱,瓶內溶液在負壓下在旋轉燒瓶內進行加熱擴散蒸發(fā)。旋轉蒸發(fā)器系統(tǒng)可以密封減壓至 400~600毫米汞柱;用加熱浴加熱蒸餾瓶中的溶劑,加熱溫度可接近該溶劑的沸點;同時還可進行旋轉,使溶劑形成薄膜,增大蒸發(fā)面積。此外,在高效冷卻器作用下,可將熱蒸氣迅速液化,加快蒸發(fā)速率。
如何選購旋轉蒸發(fā)儀
◆一 選擇旋轉蒸發(fā)器的規(guī)格
旋轉蒸發(fā)器的規(guī)格是以蒸發(fā)瓶容積區(qū)分的,所需蒸發(fā)瓶的大小,取決于被蒸發(fā)物料的數量,一般而言,2L、3L、5L旋轉蒸發(fā)器適合于實驗室小量試驗;5L、10L、20L適合于中試;20L、50L適合于中試及生產。
◆二 考慮旋轉蒸發(fā)儀蒸發(fā)速率
對同一規(guī)格的旋轉蒸發(fā)儀,影響旋轉蒸發(fā)儀蒸發(fā)速的因素主要是以下四項。
1.加熱浴溫度
加熱浴溫度愈高,物料蒸發(fā)愈劇烈。但受物料熱耐受、使用安全與能耗的制約,最常用的溫度是60℃。加熱浴溫度減去蒸發(fā)瓶傳熱溫差ΔT1和蒸發(fā)過熱度,即為物料溫度。
物料熱耐受主要指生物活性物質不發(fā)生變性的允許溫度,使用安全主要指在更換蒸發(fā)瓶的過程中無tangshang危險。加熱浴溫度適中,水的揮發(fā)速率較低,能量利用率較高。
2.旋轉蒸發(fā)儀內真空度:推薦真空度為95~98kPa(juedui壓力2~5kPa)。
3.冷凝器冷卻介質溫度:物料溫度減去冷凝器傳熱溫差ΔT2和凝結過冷度,即為所需冷凝器冷卻介質平均溫度。
物料溫度與真空度之間的關系符合物料飽和溫度與飽和壓力間的關系,在加熱浴溫度和真空度確定時,冷卻介質溫度取決于物料種類。
4.蒸發(fā)瓶轉速:蒸發(fā)瓶轉速愈高,瓶內表面浸潤面積愈大,受熱面積大;但同時液膜厚度也愈厚,增大傳熱溫差。
由于加熱浴溫度、真空度和冷卻介質溫度是關連參數,與物料種類有關,也與冷凝液回收率有關,如果需要我公司技術人員幫助,請?zhí)顚憽坝脩粜枨蟊怼薄?/span>
◆第三步 選擇配套設備
配套設備指真空設備和冷源設備。
真空泵推薦使用耐各種溶劑腐蝕的循環(huán)水真空泵或隔膜真空泵??蛇x配真空控制器,實現(xiàn)更加gaoxiao可控的蒸餾過程。
冷源設備根據冷卻介質溫度和蒸發(fā)量選擇。
注:旋轉蒸發(fā)儀的基本使用條件是:加熱浴溫度60℃、真空度95~98 kPa(壓力2~5 kPa)、冷卻介質溫度20℃。
工作參數范圍是:加熱浴溫度≤80℃、真空度≤98 kPa(壓力≥2 kPa)、冷卻介質溫度-20℃~常溫。
旋轉蒸發(fā)儀的冷凝器有立式和斜式兩種,其傳熱和流動過程不存在區(qū)別,由于立式占用空間小,一般旋轉蒸發(fā)儀采用的是立式冷凝器。